| ![]() |
Use of Ammonia Selective Ion Electrodes |
This calibration data indicates the electrodes' slopes were:
These slopes indicate the ammonia electrodes were performing properly, and were well within the normal specified response slope of 59 ± 4 mV/decade at higher ammonia concentrations.

Graphing the electrode potential (in mV) versus the log of concentration (mg/L) gives visual feedback that the electrode response is fairly linear over this concentration range.
Response time:
The ammonia electrodes' output in mV was followed over time as the electrodes were moved from one standard to another, to examine the speed of response and the time required to reach stable readings.
Response times for going from lower to higher concentration standards:

Response times for going from higher to lower concentrations are even slower.

As the above response time graphs show, upon going from lower to higher concentration the electrode provides a stable signal in about 2 to 3 minutes, with slower response at the lower end of the working range. Upon going from higher to lower concentrations, however, the gas-sensing probe requires considerably more time, on the order of 5 or 6 minutes to reach a stable signal.
The meter was set on slow filter speed, with 3 significant figures of resolution in concentration mode. The electrode signal was observed manually, and only when the electrode appeared to be stable was the standardization procedure begun on the accumet meter (allowing sufficient time for ISEs to reach equilibrium is, or should be, standard practice!). Given this time for the electrode to slowly stabilize, the Denver meter responded correctly to the electrode signal, locking on to the readings at appropriate times.
Slow response appears to be a major problem and source of error with ammonia electrodes. It is difficult to know when the electrode signal stops changing, and the result of taking readings too soon, especially with standards, is apparent drift. Running check standards frequently is critical to verifying accuracy of the electrode measurements.
Analysts report that often the readings appear to slowly drift up and then down. The downward drift can be due to loss of ammonia as gas after the ISA has been added. Use of Parafilm to seal off the beaker reduces this ammonia vaporization and helps obtain more stable readings. Providing steady stirring and allowing the electrode sufficient time to reach equilibrium (steady readings) are helpful technique issues.
Technical information provided by Denver Instrument Co.
© Fox Scientific, Inc. All rights reserved. |