Cleaning the Karl Fischer Reaction Cell |
| Type of Samples | Reagents for Cleaning |
| Hydrocarbons | Chloroform |
| Oils | Chloroform |
| Amines | Methanol |
| Ketones | Acetone |
Draw the solvent up through the frit. WARNING! Verify that the reagents that are being used for cleaning do not cause a chemical reaction with any sample material remaining in the frit!
Repeat until the frit appears to be cleaned:
More than one reagent may be drawn through the frit for cleaning purposes.
The last solvent to be drawn through the generator assembly should be either methanol or acetone.
Final cell drying should be done in an oven at 50 - 80° C for 1- 2 hours. WARNING! Let the cell cool before filling it with fresh Karl Fischer reagent, which is mostly methanol and is flammable!
After placing fresh coulometric Karl Fischer reagents in the cell and Generator Assembly, place the cell on the titrator and begin the automatic cell and reagent drying process. It may take the newly cleaned frit some time for reagent to diffuse into the frit and provide full titration current.
If the above procedure does not clean or unclog the glass frit, it may be necessary to use strong cleaning agents. For severely clogged frits, a chromic acid or concentrated nitric acid cleaning process is recommended.
WARNING! These are strong acids and must be used with appropriate care and protection!
Technical information provided by Denver Instrument Co.